Scientific version · Article 2

How Peltier cooling affects IMX477 characterisation

Comparison of two EIP campaigns on the Raspberry Pi HQ: fan cooling alone, then a thermoelectric cooler at maximum power. Protocol, measurements, statistical models and interpretative limitations.

RAW12BGGRGain 1Spatial analysis
Scope of the result. Long-exposure dark-frame temperatures change from 25–27 °C to 0–1 °C. Positive slope quantiles and several spatial metrics are lower in the cold campaign. Typical dark current remains unresolved in both conditions. This comparison alone establishes no causal thermal effect on conversion gain, read noise or the origin of the patterns.
Summary report — IMX477 · Peltier campaign6 October 2026Measurements, figures and conclusions from the EIP campaign. Read the PDF report

1. Scientific question and scope

What does moving from around 26 °C to 0–1 °C actually change in IMX477 measurements? This study extends the initial Raspberry Pi HQ characterisation. I examine general sensor properties, then the dark signatures that prevented separation of typical dark current from instrumental background shifts.

EIP handles acquisition, measurement preservation and reproducible processing. I kept the same camera in this setup for both campaigns on 6 October 2026. They compare fan cooling alone with a Peltier cooler at maximum power. The conclusions concern this experiment, not every IMX477: an observation between two campaigns does not establish causality.

2. Thermal setup and experimental protocol

From fan cooling to Peltier coupling

The warm reference was acquired without active Peltier cooling, at an ambient temperature of roughly 23 to 24 °C. Two additional fans blow towards the back of the printed circuit board (PCB) of the HQ to limit self-heating. The cooler's physical presence does not mean it is powered.

Two additional fans directed at the back of the camera's circuit board.
Warm reference: forced ventilation, with the Peltier inactive during the campaign.
Computer-aided design view of the camera layout inside PiNox.
Mechanical layout in computer-aided design (CAD). The drawing does not establish the actual state of thermal contact.

The smartphone cooler combines a thermoelectric module, a hot-side fan and electronics. The cold campaign uses the cooler at maximum power, without temperature feedback control. The module is marked FZ-2302 and measures about 25 × 25 × 4.64 mm; thickness excludes adhesive and the fan-side interface. Its exact commercial identity and electrical specifications have not been established. Terminal voltage was not measured. The marking therefore does not establish nominal voltage or current, maximum cooling power, maximum temperature difference or efficiency.

The phone contact plate is removed and the thermal paste cleaned off. A thermal pad then provides thermal coupling to the back of the PCB: heat passes through the interface and camera structure, without direct contact with the silicon. The Peltier effect moves heat to the hot side, where the integrated fan must dissipate it.

Square face of the thermoelectric module exposed after opening the cooler.
Opened module, with the FZ-2302 marking noted during disassembly.
Back of the circuit board and opened cooler before fitting the pad.
Preparing the thermal interface.
Thermal pad fitted to the back of the camera circuit board.
Pad in place before closing the thermal contact.
Open PiNox enclosure with the Raspberry Pi and cooling components.
Mechanical integration; the photograph alone does not establish the electrical operating state.

Power consumption of the complete setup

Instantaneous power is measured at the wall socket, upstream of the 230 V → USB-C power supply, with a GreenBlue GB620 E wattmeter, ASIN B0DGQLWCCN. The manufacturer specifies a range of roughly 0.3 to 3680 W and power accuracy of ±2%. This specification replaces neither independent calibration nor a complete uncertainty budget.

Wattmeter displaying 4.7 W for the Raspberry Pi alone.
Raspberry Pi alone: about 4.7 W at the socket.
Wattmeter displaying 14.0 W for the Pi and cooler at maximum power.
Pi and cooler: 14.0 W at the maximum setting used for the cold campaign.

The observed increase is about 9.3 W. It includes the Peltier module, its fan, electronics and differences in power-supply losses. Neither 9.3 W nor 14.0 W is the power of the Peltier module alone. Repeated readings, measurement stability and Pi workload are undocumented; no uncertainty for the difference or total campaign energy is calculated.

Acquisition and sampling

Both campaigns use raw data (RAW) at 12 bits, 4056 × 3040 pixels, full field. The colour filter array (CFA) is BGGR. The reference black level is 256 ADU (digital units) and the white code is 4095, leaving 3839 levels above black. The analogue gain for long-exposure dark frames is 1.

Eight exposure times are studied: 0.1, 1, 5, 10, 30, 60, 120 and 300 s. Each level contains ten images, giving 80 per campaign. Acquisition alternates exposure times, with 60 s pauses between levels to limit heating. This reduces confounding between exposure duration and heating, without constituting randomisation or independent warm/cold repetitions.

For temporal statistics, the same 16,060 positions are tracked in both campaigns; all four Bayer phases are represented. The spatial analysis in section 7 instead covers the full sensor.

The photon transfer curve uses twelve signal levels per gain, with twenty images per level, giving ten pairs. Read noise is estimated from fifty short dark frames per gain, or twenty-five pairs. Across eight gains this gives 960 photon-transfer pairs and 200 read-noise pairs, with none rejected for non-comparability. The warm reference exposure grid is set manually and the cold campaign grid automatically: illumination is not strictly identical.

3. Temperatures actually reached

Each long-exposure series has 72 available temperatures for 80 images. The first image at each level has no temperature reading; the image itself is retained. No missing value is interpolated. Ranges are 25–27 °C warm and 0–1 °C cold, with means of about 26.03 °C and 0.43 °C.

These means can have decimal places even though temperature readings are quantised to whole degrees. They summarise several integer readings; they do not improve thermometer resolution.

Mean temperatures and ranges for the eight exposure times in the warm and cold campaigns.
Long-exposure dark-frame temperatures: nine readings per level. Observed minima and maxima, not confidence intervals. The exposure-time axis is logarithmic and does not show acquisition order.

These ranges do not describe every acquisition. At gain 1, read-noise pairs cover 23–26 °C warm and 0–2 °C cold; the photon transfer curve covers 24–27 °C and 0–2 °C. Across gains, aggregated pair temperatures reach 30 °C and 3 °C.

Camera and processor temperature curves during the initial Peltier tests on 11 September.
Tests of 11 September 2026, separate from the October campaigns. Without tabulated data or a control log, no time constant or precise explanation of the changes is extracted.

4. General sensor parameters

These parameters provide context for the comparison rather than repeating the first article's full characterisation. Values remain of the same order; observed differences do not establish temperature dependence.

General properties: two campaigns, not a temperature law.
Quantity≈26 °C0–1 °CInterpretation
K at gain 1 (e⁻/ADU)2.3270962.308443Similar values; comparison conditions differ.
Read noise at gain 1 (e⁻)3.5176463.385063Same order of magnitude.
Noise at gain 22.260870 (e⁻)1.6685541.573008Observed difference; causality not established.
Behaviour across gainsContinuousContinuousNo demonstrated break across the eight gains.
Effective capacity at gain 1 (e⁻)≈8 310≈7 465Instrumental estimate tied to saturation and the fitted range.
Dynamic range at gain 1 (stops)11.20611.107Does not demonstrate a physical loss caused by cooling.

Photon transfer curve at gain 1

The photon transfer curve (PTC) relates temporal variance to mean signal in the linear range. The conversion gain K expresses electrons per ADU. Fit standard errors are ±0.001084 e⁻/ADU warm and ±0.001348 cold; they are not total instrumental uncertainties.

The coefficients of determination R² are 0.99999848 and 0.99999692 respectively. Of twelve acquired levels, nine are retained warm and eleven cold; three and one are rejected. The fits are highly linear, but their ranges and approach to saturation differ. Thermal confounding is flagged at gain 1 in the warm reference. The small difference in K remains an observation, not an established causal effect of cooling.

Read noise from pair differences

For two comparable images A and B, assuming independent fluctuations of equal variance, read noise is estimated from the standard deviation of their difference, divided by √2:

σread, ADU = σ(A − B) / √2

Conversion to electrons uses K measured at the relevant gain, not an extrapolation from gain 1 alone:

σread, e = K × σread, ADU

The units are consistent: (e⁻/ADU) × ADU = e⁻. At gain 1, digital read noise is 1.511603 ADU warm and 1.466384 cold. Pair differences separate readout dispersion from a fixed baseline, but their comparison still depends on temperatures, readout conditions and K. No uncertainty interval for the difference is available here to support attributing it to cooling.

Eight gains, no demonstrated break

Effective gains are identical: 1, 1.556231, 2.426540, 3.778598, 5.885057, 9.142858, 14.222222 and 22.260870. The selected behaviour is continuous and monotonic in both conditions. To compare a continuous model with a model containing a break, the Bayesian information criterion (BIC) accounts for both fit and model complexity.

The convention here is ΔBIC = BIC of the continuous model − BIC of the break model. Values of −8.05165 warm and −5.86966 cold therefore favour the continuous model. Statistical resampling with 2,000 repetitions gives p = 0.51524 and 0.59720. No break is statistically established across these eight gains. This does not establish the absence of a transition outside the studied range or below the protocol's sensitivity.

5. Searching for dark current

Model of the mean level

At each position i, the ten images at an exposure level are averaged. A linear regression over the eight exposure times then describes this mean level:

yi(t) = ai + mit

yi(t) is the mean level at time t, in ADU; ai is the estimated initial level, or intercept; mi is an apparent slope in ADU/pixel/s.

A positive slope can combine dark current with black-level drift, an electronic shift or a contribution common to several pixels. The slope mi therefore describes how the measured level changes, without isolating the physical current.

Methodological distinction: mean-level slope ≠ temporal-variance slope. This study fits mean level against exposure time. A future investigation will look for a signature in temporal variance. The two estimators do not measure the same thing, even though variance is already used in the PTC.

Global positive-slope criterion

The global baseline diagnostic is separate from the distribution of individual slopes. The criterion specifically looks for a positive slope greater than 1.96 times its standard error:

mglobal > 1.96 σm

σm denotes the global slope's standard error. This is the criterion used in this study, not a two-sided test for the absence of all drift.

Global diagnostic: baseline slope against exposure time.
Indicator≈26 °C0–1 °C
Slope (ADU/pixel/s)−0.00109167−0.00213482
Standard error (ADU/pixel/s)0.000921810.00095292
R²0.1894620.455478
Typical currentUnresolvedUnresolved

Both global slopes are negative. In the cold campaign, z = mglobal / σm ≈ −2.2403: this sign does not meet the search for a positive slope. The criterion therefore detects no significant positive slope, but does not test for the absence of all drift. Negative drift remains possible.

Unresolved ≠ zero. Neither campaign provides a reliable physical value for typical dark current. This result implies neither zero current, nor a negative physical current, nor a validated upper bound.

6. Slope distribution and hot pixels

The set of slopes {mi} is summarised by its median, percentiles and a robust measure of spread. Quantiles are interpolated within this measured population. p90 and p99 are lower in the cold campaign at the same positions and with the same estimator: they describe the positive tail, not a typical current. No confidence interval for their difference is available.

Comparison of apparent-slope percentiles p90 and p99 in warm and cold conditions.
Percentiles of apparent mean-level slopes against exposure time, in ADU/pixel/s. Four descriptive indicators, without uncertainty bars.
Distribution across 16,060 positions: descriptive levels and adaptive selection.
Quantity≈26 °C0–1 °C
Median (ADU/pixel/s)−0.00544253−0.00575600
p90 (ADU/pixel/s)0.009029180.00706071
p99 (ADU/pixel/s)0.040306610.03186904
Threshold (ADU/pixel/s)0.054425320.05756005
Flagged positions75 · 0.467%23 · 0.143%

The temporal hot pixels are positions whose slope exceeds a distribution-dependent threshold T:

T = max(10 × |m̃|,6 × 1.4826 × MAD)

m̃ is the median slope. The MAD (Median Absolute Deviation) is the median absolute deviation from that median: MAD = median(|mi − m̃|). The first term in the maximum sets the threshold in both campaigns.

The resulting threshold is higher in the cold campaign: about 0.05443 ADU/pixel/s warm versus 0.05756 cold. Comparing 75 with 23 is therefore not a fixed-threshold comparison. It establishes neither the elimination of 52 physical pixels nor a threefold reduction in their population.

7. Spatial structures at 300 s

For each campaign, ten 300 s images are combined using an integer temporal median into a full-sensor master at 4056 × 3040, preserving the BGGR CFA. Analogue gain is 1 and digital gain is close to 1. A low-frequency structure is then estimated on a regular grid; this derived map is not the master.

After averaging the four positions in each 2 × 2 Bayer cell, block medians estimate the low-frequency component. The grid spacing corresponds to 64 × 64 sensor pixels, or 32 × 32 Bayer cells, on a 64 × 48 block grid with partial blocks at the edges. A mask excludes values beyond percentile p99.9 and saturated values. The rejection rule is identical, but the resulting thresholds are 300 ADU warm and 282 ADU cold: the retained pixels are not exactly the same.

Warm low-frequency structure at 300 s, with visible cells on a shared greyscale.
Warm: median master of 10 RAW frames at 300 s, gain 1; block medians displayed without interpolation. Squares show the estimator's grid, not physical sensitivity regions.
Cold low-frequency structure at 300 s, with the same orientation and scale as the warm map.
Cold: median master of 10 RAW frames at 300 s, gain 1; the same orientation and greyscale range, from 254 ADU (black) to 258 ADU (white). Contrast is stretched to reveal differences.

Cells are displayed using nearest-neighbour rendering, without interpolation or antialiasing, to avoid suggesting finer spatial resolution than the estimator provides. Clipping applies only to rendering; measured values are retained.

Spatial metrics at 300 s (ADU): descriptive comparison with adaptive masks.
Metric≈26 °C0–1 °C
Edges minus centre+0.50+0.25
Corners minus centre+0.75+0.25
Row RMS0.2445330.204825
Column RMS0.0313100.022889

RMS values are rounded to six decimal places in this table. Unrounded values are 0.244532878 and 0.204825132 ADU for rows, and 0.031309741 and 0.022888862 ADU for columns. They concern the ten images from each campaign compared here, not the historical twenty-image study in the first article.

Profiles are calculated after 2 × 2 Bayer averaging and subtraction of the local block median. A median along each axis gives a row profile and a column profile, each centred on its arithmetic mean. These are spatial statistics of the master, not temporal variance.

Centred row and column profiles comparing warm and cold conditions without smoothing.
Row and column Bayer-cell profiles from median masters of 10 RAW frames at 300 s, gain 1, after removal of the low-frequency component and centring on their mean. Positions are in sensor pixels and levels in ADU; no smoothing.

For a centred profile of N values xi, the RMS (Root Mean Square) measures the typical amplitude of variations around the mean:

RMS = √[(1/N) ∑i=1N xi2]

Here it is equivalent to population standard deviation. Metrics are lower in the cold campaign, but patterns persist. Sub-ADU values result from aggregation and statistics; the quantisation of individual RAW values remains 1 ADU. These differences describe the two campaigns; they do not isolate the effect of temperature.

8. Comparability and experimental limitations

The camera, exposure times, analogue gains and tracked positions are shared by both campaigns, making comparison possible. These common elements do not isolate temperature from other conditions. Fit standard errors cannot replace an uncertainty budget covering these experimental differences. Neither a thermal law for typical current nor the origin of the patterns can be inferred.

9. What can be concluded about cooling

Observed

The cooler at maximum power allows long acquisitions at 0 to 1 °C, compared with 25 to 27 °C under fan cooling alone. The positive tail of apparent slopes is lower; several spatial metrics are reduced and fewer temporal positions are flagged as atypical, with a different threshold. Conversion gain and read noise remain of the same order. Spatial patterns remain present.

Not demonstrated

These campaigns establish neither a causal thermal law, nor a physical change in K or noise attributable to cooling, nor the patterns' origin. Typical dark current remains unresolved: it cannot be concluded to be zero. No validated upper bound is obtained. The limitation is insufficient separation of contributions to the measured signal.

What to measure next

The priority is a more sensitive, repeated study that can distinguish thermal contributions from common drift. Two campaigns are not a complete thermal characterisation.

10. Next steps

The next article will ask: “How do you measure an almost invisible dark current?” The planned protocol will compare temporal variance and pair differences, common-mode removal and separate analysis of CFA phases. It must establish repetitions and sensitivity before proposing a measurement, or an upper bound if no credible positive slope is detected. No outcome of this future study is treated as established.

Continue reading

Other version

Discovery version · Revisit the experiment through images and intuition.

In this series

  1. Article 1 · ScientificExperimental characterisation of the IMX477
  2. Article 2 · ScientificHow Peltier cooling affects IMX477 characterisation
  3. Article 3 · Coming laterHow do you measure an almost invisible dark current?

Sources consulted on 7 October 2026.